Options - VetScan i-STAT 1 Operator's Manual

Handheld analyzer
Table of Contents

Advertisement

Primary Interface
Protocol Choices
None
ASTM
Data File
AME
HL7
Check Enable external interface to enable this function.
Check Enable AutoSend to enable this function. This function automatically sends new records from the
Central Data Station to another data management system whenever they are received by the Central Data
Station. Checking this box causes the CDS program to start up with AutoSend enabled. AutoSend can also
be temporarily enabled/disabled from the CDS program. Records can also be sent manually from the Data
Viewers.
When Precision PCx glucose test strips are being run on the VetScan i-STAT 1, the glucose test strip data
can be made available to QC Manager data management program. Check the Upload PCx Strip Data to
QC Manager 2.2 box to enable this function.

Options

■ Confirmation message on exit: When this option is enabled,
a confirmation message is displayed prior to exiting the CDS
program.
■ Enable use of IR Link IDs: When enabled, the IDs pro-
grammed into the IR Links are used in determining the down-
load location of the results instead of the actual serial port or
IP address. To use this function, all IR Links must be of the
self-identifying type. This option is not recommended by
Abaxis, but is available in case the functionality is needed.
■ Accept previously deleted records: When disabled, this option prevents previously deleted
records from being stored when re-transmitted to the Central Data Station.
CDS-6
Explanation
Indicates no primary protocol in use.
Select when no external interface is used
and PCx glucose test strip data is to be
uploaded to QC Manager 3.
Data transmission conforms to ASTM
E1381-95 and ASTM E1394-97.
Formats the CDS for third-party use.
Automatic Manual Entry – Installed by
Abbott Point of Care Inc.
Data transmission conforms to HL7
(version 2.4) and is based on the CIC
Observation Reporting Interface
distributed by the National Committee
for Clinical Laboratory Science in the
USA under Document POCT-1-A. This
option is installed by Abbott Point of
Care Inc.
Central Data Station
Which Result Types
May be Transmitted
N/A
Patient Test Results ONLY
1. Test Result
2. Electronic Simulator
3. Control Result
4. Cal/Ver Result
5. Proficiency Result
1. Test Result
2. Electronic Simulator
3. Control Result
4. Cal/Ver Result
5. Proficiency Result
Patient Test Results ONLY