Inventory Log
The Inventory Log documents each action taken in the Items, Stock, Dis-
tribution and Orders tabs. Click the Date Range button in the toolbar to
select the Default date range or a Start and End date for this report. Click
the Delete button to delete entries in the log.

Customization Workspace

This workspace is used to create profiles with site specific test characteristics for the analyzers.
Customization Options and Default Settings
Option
Description
LANGUAGE
Language for text: English,
WINDOW
Japanese, German, Italian,
Dutch, Spanish, French,
Swedish, Russian, Portuguese,
Danish, and Finnish
UNIT SET
Reporting units for results.
WINDOW
Selected from predefined sets
or by analyte.
i-STAT 1
Standardization data. All non-
ANALYZER AND
expired versions listed.
CLEW WINDOWS
i-STAT 1
JAMS functionality data.
SOFTWARE
WINDOW
STATNotes
Feature allows users to
customize the Chart Page on
their i-STAT 1 Analyzers in
order to capture user-defined
information.
PREFERENCES
Options and default settings are
WINDOW
listed under seven headings:
Instrument, ID Entry, Test, QC,
Results, Analyte Enable, and
Strip Lots.
USE OPERATOR
4000 operator IDs can be stored
LIST
in the analyzer along with
certification start and end dates
for both glucose test strip and
cartridge testing.
CDS-22
Default
English
Unit Set 00
CHART0
Not enabled (no
information
stored)
Central Data Station
Comments
Russian can be downloaded only to the
Portable Clinical Analyzer. Portuguese,
Danish, and Finnish can be downloaded
only to the i-STAT 1 Analyzer.
See table below with 17 predefined unit
sets. Unit Set 99 allows the name and units
for each test to be defined individually.
Note: Reference Ranges and Action
Ranges in the Preferences Window must
be changed when changing units.
The CLEW software has an expiration
date. If an expired CLEW remains in a
customization profile, a warning will be
displayed.
Users can remotely request a JAMS
update for an i-STAT 1 Analyzer from the
CDS. For full details, see the instructions
in the software update packet, or Section 9
in the VetScan i-STAT 1 Operator's
Manual.
Contact Abaxis Technical Support for
more information on this feature.
Operator lists are created in the Operator
Workspace on the Central Data Station.
This check box cannot be enabled if the
Operator List is empty in the Operator
Workspace for all Departments (other than
Unassigned).

Hide quick links: