ABB Endura AZ30 series Programming Manual page 68

Integral probe and remote transmitter combustion oxygen monitor
Hide thumbs Also See for Endura AZ30 series:
Table of Contents

Advertisement

Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
Parameter
...Communication
Cyclic Output
Interval
Off
1, 10, 30 s
1, 10, 30, 60 m
Assignment
Oxygen
Cell Temperature
Cell Temp. Rate
Cell mV
Output
CJ Temperature
Cal. Details
Performance
Details
Mains Supply
Details
Internal
Temperature
Table 6.13 Communication Menu (Continued)
66
Comment / [Range]
For diagnostic purposes, it is possible to configure the
device (transmitter) to send out data values via the IrDA
interface on the front of the transmitter – these can be
viewed via hyperterminal.
Selects the time interval between each set of data to be
sent out via IrDA.
Selects which signals are to be transmitted.
[Assigned / Not Assigned]
6 Programming
Default
Off
Not
Assigned
(all
parameters)
COI/AZ30E–EN Rev. D