ABB Endura AZ30 series Programming Manual page 44

Integral probe and remote transmitter combustion oxygen monitor
Hide thumbs Also See for Endura AZ30 series:
Table of Contents

Advertisement

Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
Parameter
...Calibrate
AutoCal Hardware
Hardware Type
None
Internal
Remote
Test Gas Delay Time
Test Gas Detection
Test Gas 1 Detection Disabled if AutoCal Hardware / Hardware Type is set to
Disabled
Enabled
Test Gas 2 Detection Disabled if AutoCal Hardware / Hardware Type is set to
Disabled
Enabled
Table 6.7 Calibrate Menu (Continued)
42
Comment / [Range]
Selects the type of automatic calibration hardware to be
used.
Disables automatic calibration functions.
Enabled if the (optional) built-in automatic calibration is
fitted.
Enabled if an external automatic calibration system is
used.
Note. The solenoid valve outputs or relays can be used
to activate a remote automatic calibration system.
The estimated delay in seconds it takes the test gas to
pass from a remote automatic calibration unit to the
probe.
Note. For use with remote automatic calibration units
only.
[0 to 9999 s]
The probe's internal automatic calibration system has
pressure switches to detect the presence of the test
gases automatically (when enabled). This option can
also be used with remote automatic calibration units with
suitable pressure switches.
None.
If test gas is not present when a calibration is activated
an error message is created and calibration stops.
None.
If test gas is not present when a calibration is activated
an error message is created and calibration stops.
6 Programming
Default
None
0
Disabled
Disabled
COI/AZ30E–EN Rev. D