Invacare® Bora® / Spectra® XTR Series
6.
The operating hour counter is attached to the frame with a hook
and loop strip.
7.
Disconnect cable of the operating hour counter from drive
motor.
8.
Pull operating hour counter off the frame.
9.
Turn operating hour counter over.
10. Loosen and remove four Philipps head screws of the casing.
11. Disconnect plugs (+) & (-).
12.
Remove cable ties (4).
13. Carefully pull connecting cable (2) of operating hour counter out
of the casing (1).
14. Replace cable of operating hour counter.
CAUTION!
Property damage can result if the counter
is wrongly connected. If the plus and minus
wires are connected the wrong way, it will
damage the electronic components of the
operating hour counter.
– Make sure the cable is connected correctly.
15.
The cable of the operating hours counter is installed in reverse
order.
16. Connect plug on the blue wire (-) to the pin on the operating
hours counter which is marked with a "minus" symbol.
17. Connect plug on the brown wire (+) to the pin on the operating
hours counter which is marked with a "plus" symbol. The brown
wire is additionally marked with a yellow sleeve with a "plus"
symbol on it.
18. Secure cable with a cable tie (3), thereby protecting it from being
pulled off unintentionally.
19. Close casing and reattach it to the mobility device with hook
and loop fastening.
20. Install parts in reverse order.

5.9.6 Updating software

For LiNX see LiNX Service Manual. For VR2, R-Net, Shark,
ACS2 see revision 15 of this service manual.
30

5.9.7 Batteries

CAUTION!
Injury hazard and possible material damages if
batteries are handled improperly
The installation of new batteries may only be carried out
by authorised specialists.
– Observe the warning information on the batteries.
– Only use battery versions stated in the specifications.
CAUTION!
Fire and burns hazard if battery terminal is
bypassed
– Take great care to ensure that the battery terminals
are never short-circuited with tools or mechanical
mobility device parts.
– Ensure that the battery terminal caps have been
replaced if you are not working on the battery
terminals.
CAUTION!
Danger of crushing
The batteries are extremely heavy. This results in injury
hazards to your hands. Bear in mind that the batteries
are sometimes very heavy.
– Handle the batteries with care.
WARNING!
Burn hazard
Injury hazard due to discharged acid.
– Always wear acid-proof protective gloves when
handling batteries.
– Always wear protective goggles when handling
batteries.
What to do if acid is discharged
– Always take clothing which has been soiled by or
dipped in acid off immediately!
– Rinse any areas of your skin which has come into
contact with battery acid off immediately with plenty
of water!
If contact with eyes is made
– You should also consult an eye specialist immediately
afterwards!
When disassembling, take care of small parts such as screws
and washers. Put all small parts down so that they can be
reassembled in the right sequence.
Removing battery boxes
CAUTION!
Risk of crushing
The batteries are extremely heavy. This results in injury
hazards to your hands.
– Bear in mind that the batteries are sometimes very
heavy.
– Handle batteries with care.
A spare fuse is located behind the Invacare logo on the rear
battery.