Maintenance Interval; Checking The Records - Sirona galileos Maintenance Manual

Hide thumbs Also See for galileos:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
5.2

Maintenance interval

NOTE
In addition to the yearly inspection carried out by the system owner, safety-related checks must be
performed by the service engineer after 4, 7 and 10 years, and then at two-year intervals.
5.3

Checking the records

In the chapter
"Annual inspection performed by the system owner or other authorized persons" on page 6
in the chapter "Yearly calibration by the service engineer (for the USA, see the Maintenance Instruc-
tions)" on page 15 and in the chapter
IEC 62353" on page 17.
Checking the records of the yearly
inspection by the system owner
Maintenance interval
(after years)
Date
(enter month/year)
Annual inspection performed
by the system owner
(see Chapter 2)?
Yearly calibration has been per-
formed by the service engineer
(see Chapter 3 or Maintenance
Instructions)?
Have all safety-related checks
been carried out and did the mea-
sured values remain constant?
Records complete?
All questions answered with yes?
All work required due to a nega-
tive result during the checks has
been performed?
61 25 533 D3437
D3437.105.01.10.02
"Safety-related checks by the service engineer in compliance with standard
Answer questions with yes () or no (−)
4
7
10
If no  Perform annual inspection and calibration prior to maintenance.
5 Maintenance by the service engineer