Troubleshooting; Provider's Notes - DeVilbiss Vacu-Aide 7314 Series Instruction Manual

Qsu suction unit aspirator
Hide thumbs Also See for Vacu-Aide 7314 Series:
Table of Contents

Advertisement

6' Patient Tubing
1
2
Disconnect from lid.
Rinse thoroughly by running
warm tap water through it.
Suction Unit
1
2
Disconnect from power source. Wipe the housing with a clean
Turn unit
"off" and allow
vacuum to drop.
CAUTION– Do not submerge in water as this will result in damage to the vacuum pump.

TROUbLESHOOTING

DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a qualified
DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
NOTE– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO YOUR DEVIlBISS
HEAlTHCARE PROVIDER WHERE YOU PURCHASED IT, PlEASE TAkE A FEW mOmENTS TO CHECk FOR THESE POSSIBlE CAUSES:
PRObLEM
Unit does not turn on when external power is connected.
Green external power light does not illuminate.
Pump runs, but there is no suction.
Low suction.
Unit does not turn on (no external power is connected).
7314P Series Only
Battery will not charge (external power and charge
indicator lights should be illuminated during charge
mode) 7314P Series Only

PROVIDER'S NOTES

No routine calibration or service is required provided the device is used in accordance with the manufacturer's directions. In case of a change of patient, the device must be
reconditioned to protect the user. Reconditioning must only be carried out by the manufacturer or service provider. Between patients:
1.
Visually inspect unit for any damage, missing parts etc.
2.
Ensure that unit and accessories are clean.
3.
Use an independent vacuum gauge to verify the unit provides the proper vacuum level as stated in Specifications.
4.
Discard and replace collection container, filter, and tubing between patients.
A-7314
3
4
Follow by soaking in a solution
Keep outer surface clean by
of 1 part vinegar (>=5% acetic
wiping with clean, damp cloth .
acid concentration) to 3 parts
water (131ºF-149ºF/55ºC-
65ºC) for 60 minutes.Rinse
with clean, warm water and air
dry.
3
cloth and any commercial
(bacterial-germicidal)
disinfectant.
ACTION
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded.
4. Check for leaks or cracks in collection container assembly.
1. Use vacuum regulator knob to increase suction level.
2. Check system for leaks.
1. Check that battery is fully charged and/or charge battery.
1. Verify that both external power and charge indicator lights illuminate.
2. Check power sources and connections.
3. Ensure wall outlet is live by plugging in a lamp.
AC to DC Adapter
1
2
Disconnect AC to DC adapter
Wipe AC to DC adapter
from unit and from power
housing and cords with a dry
source.
cloth.
Carrying Case
1
Wipe with clean cloth
dampened with detergent or
disinfectant.
EN - 7