9.4 Error Displays
The display appears as follows when an error occurs on the logger.
Logger Error Displays
Error Displays
or
LR5000 Utility Program Error Displays
Error Displays
OF
UF
Meaning
Calibration data error:
A fault occurred with the internal
calibration data.
Microcomputer error:
A fault occurred in microcomputer
ROM/RAM.
Data recording error:
A fault occurred in recording data
or accessing settings.
Battery voltage is too low for nor-
mal logger operation.
Measurement
stopped
recording because the battery is
depleted.
A measured value is out of range.
• The sensor plug is inserted
incorrectly.
• The sensor plug is not inserted
all the way in.
• The sensor is damaged.
• The logger is damaged.
Meaning
A measured value is out of range.
9.4 Error Displays
Remedies and References
Inspection and repair is required.
Please contact your dealer or Hioki
representative.
See:
"Requesting repairs" (p.91)
Replace the battery.
See:
"2.1 Installing (or Replacing)
the Battery" (p.17)
Replace the battery.
Resume recording after battery re-
during
placement.
See:
"2.1 Installing (or Replacing)
the Battery" (p.17)
Out-of-range values cannot be dis-
played.
If the humidity display indicated O.F.,
dry the sensor and try again, as there
may be moisture on the sensor.
[OF] or [UF] is displayed when this
data is imported by the LR5000 Utility
Program.
Verify the correct plug orientation,
and insert it as far as possible.
If the values are not displayed despite
these measures, the sensor and log-
ger need to be inspected and repaired.
Please contact your dealer or Hioki
representative.
See:
"Requesting repairs" (p.91)
[ERROR]
is displayed when this
(faulty) data is imported by the
LR5000 Utility Program.
Remedies and References
Out-of-range values cannot be dis-
played.
95
3
9