Minimum Output .................................................... 9
Inductive Load Limits ........................................... 400
Power Factors ........................................................ 5730A, 0.9 to 1; 5725A, 0.1 to 1. Subject to compliance voltage
Frequency
Range (Hz) .......................................................... 10.000 - 11.999, 12.00 - 119.99, 120.0 - 1199.9, 1.200 k - 10.000 k
Specification .................................................. ±0.01 % of output
Resolution ..................................................... 11,999 counts
Settling Time ...................................................... 5 seconds for 5730A ranges; 6 seconds for 5725A 11 A range; +1
Overshoot .......................................................... <10 %
Wideband AC Voltage (Option 5730A/03 and 5730A/05) Specifications (99 %
Confidence Level)
Specifications apply to the end of the cable and 50 Ω termination used for calibration.
Volts
1.1 mV
3.3 mV
11 mV
33 mV
110 mV
330 mV
1.1 V
3.5 V
Frequency
(Hz)
10 - 30
30 - 119.99
120 - 1.1999 k
1.2 k - 11.999 k
12 k - 119.99 k
120 k - 1.1999 M
1.2 M - 2 M
2 M - 11.9 M
12 M - 20 M
20 M - 30 M
30 M - 50 M
Note:
1.
For output voltages <50 % of full range in the 33 mV, 110 mV, 330 mV, 1.1 V , and 3.5 V ranges, add 0.1 % to the amplitude
flatness specification.
Additional Operating Information:
dBm reference = 50 Ω
Range boundaries are at voltage points, dBm levels are approximate.
dBm = 10 log (
2.
Applies to Option 5730A/05 only.
40
Range
dBm
-46
-37
-26
-17
-6.2
+3.4
+14
+24
Amplitude Flatness, 1 kHz Reference
Frequency
Resolution
1.1 mV
(Hz)
±(% output + floor indicated)
0.01
0.3
0.01
0.1
0.1
0.1
1
0.1
10
0.1
100
0.2 + 3 μV
[1]
1 k
0.2 + 3 μV
1 k
0.4 + 3 μV
10 k
0.6 + 3 μV
10 k
1.5 + 15 μV
[2]
10 k
3.0 + 15 μV
Power
) ; 0.22361 V across 50 Ω = 1 mW or 0 dBm
1
mW
for 220
range, 10 % on all other ranges. 1 A minimum for
μA
μA
5725A.
(5730A, or 5725A). 20
μH
limits.
second for amplitude or frequency range change.
Absolute / ±5 °C from calibration temperature
Resolution
24 Hours
10 nV
0.4 + 0.4
10 nV
0.4 + 1
100 nV
0.2 + 4
100 nV
0.2 + 10
0.2 + 40
1 μV
1 μV
0.2 + 100
10 μV
0.2 + 400
0.15 + 500
10 μV
Voltage Range
3.3 mV
>3.3 mV
0.3
0.3
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1 + 3 μV
0.1 + 3 μV
0.1 + 3 μV
0.1 + 3 μV
0.3 + 3 μV
0.2 + 3 μV
0.5 + 3 μV
0.4 + 3 μV
1.5 + 3 μV
1 + 3 μV
3.0 + 3 μV
2.0 + 3 μV
for 5730A output >1 A.
μH
30 Hz - 500 kHz
90 Days
180 Days
±(% output + μV)
0.5 + 0.4
0.6 + 0.4
0.45 + 1
0.5 + 1
0.35 + 4
0.5 + 4
0.3 + 10
0.45 + 10
0.3 + 40
0.45 + 40
0.25 + 100
0.35 + 100
0.25 + 400
0.35 + 400
0.2 + 500
0.3 + 500
Settling Time
Temperature
To Published
Coefficient
Specification
±ppm/°C
(Seconds)
100
7
100
7
100
5
100
5
100
5
100
5
100
0.5
100
0.5
150
0.5
300
0.5
600
0.5
1 Year
0.8 + 2
0.7 + 3
0.7 + 8
0.6 + 16
0.6 + 40
0.5 + 100
0.5 + 400
0.4 + 500
Harmonic
Distortion
(dB)
-40
-40
-40
-40
-40
-40
-40
-40
-34
-34