Minimum Output ................................................... 300 μV (-57 dBm)
VSWR at Output Terminal .................................... <1.1 Typical
Frequency Specification ...................................... ±0.01 % of output
Frequency Resolution .......................................... 11,999 counts to 1.1999 MHz, 10799 counts to 11.999 MHz, 3800
Overload Protection .............................................. A short circuit on the wideband output will not result in damage. After

52120A Specifications when Operated with the 5730A

Line Power
Voltage range ..................................................... 100 V to 240 V
Frequency ........................................................... 47 to 63 Hz
Voltage variations ............................................... ±10 % about line voltage
Power consumption ............................................ <1500 VA
Dimensions (HxWxL)
With feet .............................................................. 192 mm x 432 mm x 645 mm (7.6 in x 17.0 in x 25.5 in)
Without feet ......................................................... 178 mm x 432 mm x 645 mm (7.0 in x 17.0 in x 25.5 in)
Weight .................................................................... 25 kg (54 lb)
Temperature
Operating ............................................................ 5 °C to 35 °C (41 °F to 95 °F)
Calibration (tcal) .................................................. 16 °C to 30 °C (61 °F to 86 °F)
Storage ............................................................... 0 °C to 50 °C (32 °F to 122 °F)
Transit ................................................................. -20 °C to +60 °C (-4 °F to +140 °F) <100 hours
Warmup Time ........................................................ Twice the time since last warmed up, to a maximum of 1 hour.
Humidity (non-condensing)
Operating ............................................................ <80 %, 5 °C to 31 °C (41 °F to 88 °F) ramping linearly down to 50 % at
Storage ............................................................... <95 %, 0 to 50 °C (32 °F to 122 °F)
Altitude
Operating ............................................................ 2,500 m (8,200 ft) maximum
Non-Operating .................................................... 12,000 m (39,400 ft) maximum
Shock and Vibration ............................................. MIL-PRF-28800F Class 3
Safety ..................................................................... EN/IEC 61010-1, 300 V CAT II, Pollution Degree 2
Electromagnetic Environment ............................. IEC 61326-1, Industrial
Electromagnetic Compatibility ............................ FCC Rules part 15 sub part B
Indoor use only ..................................................... IP20
counts to 50 MHz
settling time, normal operation is restored upon removal.
35 °C (95 °F)
Applies to use in Korea only. Class A Equipment (Industrial
Broadcasting & Communication Equipment)
[1]
This product meets requirements for industrial (Class A)
electromagnetic wave equipment and the seller or user should
take notice of it. This equipment is intended for use in business
environments and is not to be used in homes.