Technical Specifications - Seitron CHEMIST 400 Use And Maintenance Manual

Combustion analyzer
Table of Contents

Advertisement

2.0

TECHNICAL SPECIFICATIONS

2.1 Technical Specifications
Autozero:
Dilution:
Self-diagnosis:
Type of combustible:
Power:
Battery charger:
Charging time:
Instrument working time:
Printer:
Printer powered:
Printer autonomy:
Internal data memory:
User data:
Print-out heading:
Display:
Communication port:
Bluetooth (optional):
Line filter:
Suction pump:
Condensate trap:
Carbon black:
Leak test:
Condensing boiler efficiency:
Environmental gases:
Draught test:
Operating temperature range: -5°C to +45°C
Storage temperature range:
Operating humidity range:
Protection grade:
Air pressure:
Outer dimensions: Analyser: 30,7 x 10,5 x 9,6 cm (W x H x D)
Case:
Weight:
Analyser: ~ 1.1 kg
Compliant with the European standard EN50379-1 and EN50379-2 for the following measurements:
O
2
CO medium
NO
Temperature (flue gas)
Temperature (supply air)
Pressure (draft)
Pressure (differential)
K900000000SE 022628A0 071013
Automatic autozero cycle (with probe inserted in the stack except for Chemist
400B and Chemist 401 versions).
Expansion system of the CO sensor measuring range up to 50,000ppm
(5.00%v/v) programmable as a simple protection of the CO sensor with triggering
threshold programmable by the user. Preset triggering threshold at 1,500 ppm.
All the functions and internal functions are checked and anomalies signalled.
7 predefined by the factory and 16 that can be programmed by the user.
Li-Ion battery pack with internal protection circuit.
External battery charger.
2 hours for charging from 0% to 90% (3 hours for 100% charge).
18 hours of non-stop operation (excluding printing).
Internal, 24-column impact printer, using ordinary paper (roll 18 m long and
57 mm wide).
By the analyser batteries.
Up to 40 analysis reports with the batteries fully charged.
300 complete data analyses, time and name of the customer can be stored.
3 programmable user names.
4 lines x 24 characters, customisable by the user.
Graphic backlit LCD, measuring 42 x 60 mm.
USB with mini-USB connector.
Communication range: <100 meters (free field)
Manufacturer: FREE2MOVE
Model: F2M03GLA - CLASS 1
Bluetooth SIG Qualification Design (QDL) Certificate: B012541
Certification EC-R & TTE:
With replaceable cartridge, 99% efficient with 20um particles.
1.2 l/min heads at the flue up to 135mbar.
Outside the instrument.
Using an external hand pump; it is possible to enter and print the Bacharach
index.
Gas pipes tested for leaks with separate printout of the result, by means of the
attachment AACKT02, according to UNI 7129 (new systems) and UNI 11137:
2012 (existing systems), with automatic calculation of pipe volume.
Automatic recognition of the condensing boiler, with calculation and printout of
efficiency (>100%) on the LHV (Lower Heating Value) in accordance with
UNI10389-1.
Measurement and separate printout of the ambient CO and NO values.
Draught tested as per the UNI 10845 standard. Using the external draught gauge
AACDP02 the resolution is 0.1 Pa with 0.5 Pa accuracy.
-20°C to +50°C
20% to 80% RH
IP42
Atmospheric
48.2 x 37.5 x 16 cm (W x H x D)
0681
1
2
3
4
5
6
7
8
9

Advertisement

Table of Contents
loading

Table of Contents