Flow Chart - Info Menu - Seitron CHEMIST 400 Use And Maintenance Manual

Combustion analyzer
Table of Contents

Advertisement

4.8.1 Flow Chart - Info Menu

Activates the Info Menu.
INFORMATION
BATTERY STATUS
►Battery status
Bat:
Sensors Config.
Vbat:7.82 V
Sensors diagnostic
Gas Path Check
Memories diagnostic
INFORMATION
Battery status
►Sensors Config.
Sensors diagnostic
Gas Path Check
Memories diagnostic
QUIT
INFORMATION
SENSOR DIAGNOSTIC
Battery status
►1:O
Sensors Config.
2:CO
►Sensors diagnostic
3:NO
Gas Path Check
4:SO
Memories diagnostic
SENSOR DIAGNOSTIC
1:O
►2:CO
3:NO
4:SO
SENSOR DIAGNOSTIC
1:O
2:CO
►3:NO
4:SO
SENSOR DIAGNOSTIC
1:O
2:CO
3:NO
►4:SO
K900000000SE 022628A0 071013
91 ٪
SENSORS TYPE
O
NO
2
1 3
2 4
CO
SO
2
ok
2
ok
ok
ok
2
ok
2
ok
ok
ok
2
ok
2
ok
ok
ok
2
ok
2
ok
ok
ok
2
To return to the previous screen,
press
The battery symbol filling up means that battery charging is
ongoing.
This screen page shows, for each position, the following
messages (example referred to the sensor in position 3):
NO
Sensor configured OK
□→NO
Sensor missing or non communicating
NO→□
New sensor detected
CO
Sensor detected in a wrong position
NO→NO
A sensor different from the previous one has
2
been detected.
Use arrows
to scroll the parameters of each selected
cell. Here below are the data which can be displayed
through the sensors troubleshooting menu:
Type:
Type of sensor
Revision:
Sensor revision index
Date code:
Production batch
Gas:
Gas measured
Serial:
Sensor serial number
Manufacturing date:
Production date
Calibration date:
Calibration date
Is:
Sensor Is current
Ia:
Sensor Ia current
Furthermore, under sensor troubleshooting mode, the
instrument can display the status of each single cell (here is
an example) visualized after the visualization of the
measured gas:
Ok:
No problem detected
missing
The sensor has not been detected
data err
Sensor memory data error
Unknown
The instrument FW needs to be
updated
pos err
Sensor
position
cal err
Calibration error
curr err
Currents out of range
non config
This sensor is not to be used
because it has not been accepted in
the page 'type of sensor'.
.
installed
in
the
wrong
21
1
2
3
4
5
6
7
8