Unit Calibration - Dräger Polytron 3000 Instructions For Use Manual

(approved as type p3s). transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 3000:
Table of Contents

Advertisement

Maintenance

Unit calibration

Caution:
— When the transmitter is installed in Ex areas zone 22 or Class II, Div. 1 & 2,
Group E, F, G the opening of the housing (required for calibration) must not
be done when connected to power (power must be turned off or the area
has to be declassified). Explosion hazard!
— Ensure that the sensor is warmed up before it is calibrated. See the sensor data
sheet for the warming-up time.
— Only the zero point is checked if an oxygen sensor has been fitted. The zero point
of an oxygen sensor does not require calibration.
— The transmitter can be calibrated by the operator on site.
— For critical applications, the calibration intervals should be defined in accord-
ance with the recommendations in EN 50073
lations.
Note the calibration sequence!
First check the zero point and correct it necessary, Immediately after this, check
the sensitivity and adjust it as necessary.
— Never calibrate the sensitivity before calibrating the zero point.
— Zero gas and test gas: see the information in the sensor data sheet.
Caution:
Test gas must not be inhaled. Risk to health! Care must be taken about the
risks which can arise when using test gas; hazard instructions and safety advice
must be observed.
For details, see appropriate DIN Safety Data Sheets.
Open the front cover of the service port with an Allen key by turning anticlockwise
o
(approx. 60
). The maintenance switch and potentiometers for calibration are now
revealed.
Attention!
Use only a 5 mm Allen key without a ball head.
Note:
The Dräger Polytron 3000 does not support the storage of calibration data
in the sensor data base.
Measuring / maintenance mode
1 Maintenance switch with two positions.
2 Measuring mode position (left-hand position) – measured values are relayed to
the analogue output.
3 Maintenance mode position (right-hand position) – a maintenance signal (3.4 mA
±0.2 mA constant) is relayed to the analogue output and prevents alarms being
triggered.
1)
EN 50073 – Guidelines for selection, installation, use and maintenance of devices for the
detection and measurement of flammable gases and oxygen.
2)
EN 45544-4 – Electrical devices for the direct detection and direct concentration measure-
ment of toxic gases and vapours – Part 4: Guidelines for selection, installation, use and
maintenance.
14
1)
2)
, EN45544-4
and national regu-
Polytron
1
2
3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents