+20 dB Range Flatness Adjustment

Caution
Most power meters will require an attenuator (–20 dB) or special power
head to measure the +20 dB output.
Be sure to correct the measurements for the specifications of the
attenuator you use. For example, if the nominal attenuator value
is –20 dB at the specified frequency, you must add 20 dB to the power
meter reading before entering the value.
1 Make the connections as shown below:
2 Use the power meter to measure the output amplitude for each of the
setups in the table on the next page.
Note
Setup 63 establishes the power meter reference for all the remaining
setups in this table. You must always perform setup 63 before any of the
following setups.
Chapter 4 Calibration Procedures
+20 dB Range Flatness Adjustment