Self-Test - Keysight 33210A Service Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

Note
4
Note
Chapter 4 Calibration Procedures

Self-Test

Self-Test
Self-Test is performed as the first step to ensure the instrument is in
working order before beginning any additional adjustments.
Be sure to follow the requirements listed in "Test Considerations" on
page 57 before beginning any adjustments.
1 Press
on the front panel. Select Perform Cal on the "Test / Cal"
menu. Enter setup number "1" and select BEGIN.
Setup
1
Performs the Self-test. The Main Output is disabled during test.
2 If the instrument fails any self-test, you must repair the instrument
before continuing the adjustment procedures.
The self-test procedure takes approximately 15 seconds to complete.
76