Chapter 4 Calibration Procedures

Automating Calibration Procedures

Automating Calibration Procedures
You can automate the complete verification and adjustment procedures
outlined in this chapter if you have access to programmable test
equipment. You can program the instrument configurations specified
for each test over the remote interface. You can then enter read-back
verification data into a test program and compare the results to the
4
appropriate test limit values.
You can also adjust the instrument from the remote interface. Remote
adjustment is similar to the local front-panel procedure. You can use a
computer to perform the adjustment by first selecting the required
function and range. The calibration value is sent to the instrument and
then the calibration is initiated over the remote interface. The instrument
must be unsecured prior to initiating the calibration procedure.
4
For further information on programming the instrument, see chapters
3 and 4 in the Keysight 33210A User's Guide.
55