Frequency (Internal Timebase) Adjustment

The function generator stores a calibration constant that sets the VCXO
to output exactly 10 MHz.
1 Set the frequency counter resolution to better than 0.1 ppm and the
input termination to 50 Ω (if your frequency counter does not have a 50 Ω
input termination, you must provide an external termination). Make the
connections shown below.
2 Use a frequency counter to measure the output frequency for each setup
in the following table.
Nominal Signal
Setup
Frequency
2
<10 MHz
3
>10 MHz
4
~10 MHz
5*
10 MHz
* Constants are stored after completing this setup.
3 Using the numerical keypad or knob, adjust the displayed frequency at
each setup to match the measured frequency. Select ENTER VALUE.
4 After performing setup 5:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform "Internal
Timebase Verification", on page 62.
b. If you are making all the adjustments and then verifying the
instrument's performance, continue with the next procedure in
this chapter.
Chapter 4 Calibration Procedures
Frequency (Internal Timebase) Adjustment
Amplitude
1 Vpp
Output frequency is slightly less than 10MHz
1 Vpp
Output frequency is slightly more than 10MHz
1 Vpp
Output frequency should be near 10MHz
1 Vpp
Output frequency should be 10MHz ±1ppm
4
4
77