0 dB Range Flatness Adjustments

1 Connect the power meter as shown on page 87.
2 Use the power meter to measure the output amplitude for each of the
setups in the table below.
Note
Setup 39 establishes the power meter reference for all the remaining
setups in this table. You must always perform setup 39 before any of the
following setups.
Setup
Frequency
39*
100 kHz
40*
200 kHz
41*
500 kHz
42*
1.5 MHz
43*
3 MHz
44*
4 MHz
45*
6 MHz
46*
8 MHz
47*
9.9 MHz
48*
200 kHz
49*
500 kHz
50*
1.5 MHz
51*
3.0 MHz
52
53
* Constants are stored after completing this setup.
Nominal Signal
Amplitude
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
0.28 Vrms
2 dBm
Chapter 4 Calibration Procedures
0 dB Range Flatness Adjustments
Power Meter Reference for 0dB Range
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Elliptical Filter
Flatness for 0 dB, Linear Phase Filter
Flatness for 0 dB, Linear Phase Filter
Flatness for 0 dB, Linear Phase Filter
Flatness for 0 dB, Linear Phase Filter
Setup not used for this instrument
Setup not used for this instrument
4
4
85