Chapter 4 Calibration Procedures

Recommended Test Equipment

Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not
available, substitute calibration standards of equivalent accuracy.
Requirements
ac volts, true rms, ac coupled
accuracy: ±0.02% to 1 MHz
dc volts
accuracy: 50 ppm
resolution: 100 µV
Resistance
Offset-compensated
accuracy: ±0.1Ω
100 kHz to 100 MHz
1 µW to 100 mW (–30 dBm to +20 dBm)
accuracy: 0.02 dB
resolution: 0.01 dB
100 kHz to 100 MHz
1 µW to 100 mW (–30 dBm to +20 dBm)
–20 dB
accuracy: 0.1 ppm
500 MHz
2 Gs/second
50Ω input termination
N type (m) to BNC (m)
BNC (m) to dual-banana (f)
Dual banana (m) to dual banana (m)
RG58, BNC (m) to dual banana
RG58, BNC (m) to BNC (m)
* Q = Quick Verification
** An oscilloscope is not required for calibration, only for troubleshooting.
56
Recommended Model
Keysight 3458A
Keysight E4418B
Keysight 8482A
Keysight 8491A Opt 020
Keysight 53131A Opt
010 (high stability)
Keysight 54831B
N type (m) to BNC (m)
Keysight 10110B
Keysight 11000-60001
Keysight 11001-60001
Keysight 8120-1840
P = Performance Verification
Use*
Q, P, T
Q, P, T
Q, P, T
Q, P, T
Q, P, T
T
Q, P, T
Q, P, T
Q, P, T
Q, P, T
Q, P, T
T = Troubleshooting