Ac Amplitude (High-Impedance) Verification - Keysight 33210A Service Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

AC Amplitude (high-impedance) Verification

This procedure checks the ac amplitude output accuracy at a frequency
of 1 kHz, and establishes reference measurements for the higher
frequency flatness verification procedures.
1 Set the DMM to measure Vrms Volts. Connect the DMM as shown below.
2 Set the instrument to each output described in the table below and
measure the output voltage with the DMM. Press
impedance to High–Z. Be sure the output is enabled.
Output Setup
Function
Q
High Z
Sine Wave
Q
High Z
Sine Wave
Q
High Z
Sine Wave
Q
High Z
Sine Wave
Q
High Z
Sine Wave
Q
High Z
Sine Wave
Q
High Z
Square Wave
1
2
3
4
3 Compare the measured voltage to the test limits shown in the table.
Keysight 33210A
Frequency
1.000 kHz
1.000 kHz
1.000 kHz
1.000 kHz
1.000 kHz
1.000 kHz
4
1.000 kHz
Enter the measured value on the worksheet (page 61) as 1kHz_0dB_reference.
Enter the measured value on the worksheet (page 61) as 1kHz_10dB_reference.
Enter the measured value on the worksheet (page 61) as 1kHz_20dB_reference.
Based upon 1% of setting ±1 mVpp (50 Ω); converted to Vrms for High–Z.
Chapter 4 Calibration Procedures
AC Amplitude (high-impedance) Verification
Amplitude
Nominal
20.0 mVrms
0.020 Vrms
67.0 mVrms
0.067 Vrms
200.0 mVrms
0.200 Vrms
670.0 mVrms
0.670 Vrms
2.000 Vrms
2.0000 Vrms
7.000 Vrms
7.000 Vrms
900.0 mVrms
0.900 Vrms
4
to set the output
Measurement
4
Error
± 0.00091 Vrms
± 0.00138 Vrms
± 0.00271 Vrms
1
± 0.00741 Vrms
2
± 0.0207 Vrms
3
± 0.0707 Vrms
± 0.0100 Vrms
63