Performance Verification Tests - Keysight 33210A Service Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

4
Chapter 4 Calibration Procedures

Performance Verification Tests

Performance Verification Tests
Use the Performance Verification Tests to verify the measurement
performance of the instrument. The performance verification tests use
the instrument's specifications listed in the "Specifications" chapter
beginning on page 11.
You can perform three different levels of performance verification tests:
• Self-Test A series of internal verification tests that give high
confidence that the instrument is operational.
• Quick Verification A combination of the internal self-tests and
selected verification tests.
• Performance Verification Tests An extensive set of tests that are
recommended as an acceptance test when you first receive the
instrument or after performing adjustments.
Self-Test
A brief power-on self-test occurs automatically whenever you turn on the
instrument. This limited test assures that the instrument is operational.
To perform a complete self-test:
1 Press
on the front panel.
2 Select the Self Test softkey from the "Test/Cal" menu.
A complete description of the self-tests can be found in chapter 6.
The instrument will automatically perform the complete self-test
procedure when you release the key. The self-test will complete in
approximately 15 seconds.
• If the self-test is successful, "Self Test Passed" is displayed on the
front panel.
• If the self-test fails, "Self Test Failed" and an error number are
displayed.
• If repair is required, see chapter 6," Disassembly and Repair," for
further details.