Db Range Flatness Adjustments - Keysight 33210A Service Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

+10 dB Range Flatness Adjustments

Note
The Linear Phase path is not adjusted. It is approximated using the
other path's values.
1 Connect the power meter as shown below.
2 Use a power meter to measure the output amplitude for each of the
setups in the table on the next page.
Note
Setup 54 establishes the power meter reference for all the remaining
setups in this table. You must always perform setup 54 before any of the
following setups.
Chapter 4 Calibration Procedures
+10 dB Range Flatness Adjustments
4
4
87

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents