Chapter 4 Calibration Procedures

Time Required for Calibration

Time Required for Calibration
The Keysight 33210A can be automatically calibrated under computer
control. With computer control you can perform the complete calibration
procedure and performance verification tests in approximately 30 minutes
once the instrument is warmed-up (see "Test Considerations" on page 57).
Manual adjustments and verifications, using the recommended test
equipment, will take approximately 2 hours.
START
Incoming
Verification?
NO
Perform
Adjustments
(approx 1 Hour)
Do Performance
Verification Tests
(approx 1 Hour)
DONE
54
YES
Do Performance
Verification Tests
(approx 1 Hour)