2.2 Operating Page
2.2.1 Single PID Controller
Note. The Single PID Controller operating page replaces the Single Input pH operating page shown in Section 2.3.1 of the User
Guide, IM/AX4PH.
7. 0 0
pH
24. 6
Deg.C
Setpoint 7.00pH
7. 0 0
pH
----
Manual
Control Mode
7. 0 0
pH
60. 0
% Man
Setpoint 7.00pH
7. 0 0
pH
7. 0 0
pH
Control Setpoint
404
mV
Millivolts
100. 0
%
7. 0 0
pH
A: Slope & Check
Measured Values
pH.
Temperature.
Control set point.
Control Mode
Measured pH.
Control mode.
Use the
and
keys to switch between manual (
Control Output
Measured pH.
Control output (%), manual (
When
is set to
Control Mode
control output between 0 and 100%.
Control Set Point
Measured pH.
Control set point.
Use the
and
keys to adjust the control set point between 0 and 14pH.
Measured Millivolts
Millivolts.
% Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see
page – User Guide Section 5.3) and 105% is displayed. If the value is
CONFIG. SENSORS
outside these limits, check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7pH is the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Alarm 3 set to
Wash Function
Section 2.3.3.
Alarm 3 not set to
Monitoring pH
Manual
) or automatic (
).
Man
Auto
(see above), use the
Manual
(
, User Guide Section 5.4) – see User Guide
Wash
A3: Type
.
Wash
2 OPERATION...
) and automatic (
) control.
Auto
and
keys to adjust the
in the
Set Min Slope
7