...PROGRAMMING
...4.3 Configure Control

...4.3.1 Configure Single PID Controller

Output Type
set to
Time
10. 0
Secs
Cycle Time
PID Controller
Output Type
set to
Analog
4-20mA
0-20mA
----
0-10mA
Output Range
PID Controller
14
Time Proportioning Output
The time proportioning output value is calculated using the following equation:
control output x cycle time
on time =
100
Set the cycle time between 1.0 and 300.0 seconds in 0.1 second increments – see Fig. 4.1
Mode C.
Note. Changes to the cycle time do not take effect until the start of a new cycle.
Output = 0%
Energized
Output = 25%
De-energized
Energized
Output = 50%
De-energized
Energized
Output = 75%
De-energized
Output = 100%
See Section 4.4.
Power Recovery
Optional analog output board fitted – see User Guide, Section 5.6.
CONFIG. CLOCK
Optional analog output board not fitted – see User Guide, Section 5.7.
CONFIG. SECURITY
Analog Output
Set the analog current output range.
See Section 4.4.
Power Recovery
Optional analog output board fitted – see User Guide, Section 5.6.
CONFIG. CLOCK
Optional analog output board not fitted – see User Guide, Section 5.7.
CONFIG. SECURITY
Permanently de-energized
2.5s
7.5s
5s
7.5s
Permanently energized
Cycle Time = 10s
5s
2.5s