Configure Power Failure Recovery Mode - ABB AnalyzeIT AX460 User Manual

Single input ph/redox (orp) analyzer
Hide thumbs Also See for AnalyzeIT AX460:
Table of Contents

Advertisement

4.4 Configure Power Failure Recovery Mode

----
Power Recovery
Last
Manual
----
Auto
Power Rec. Mode
50. 0
%
Default Output
Power Recovery
Power Failure Recovery Mode
When power to the analyzer is restored,
according to the chosen Power Failure Recovery Mode.
Set the required mode:
Auto
Control Mode
failure.
Manual –
Control Mode
failure.
Control Output
frame below.
Output
Last
Control Mode
to the power failure.
Default Output
If
is set to
(see Section 4.3), set the default output required between 0 and
Controller
PID
100% in 0.1% increments.
If
is set to
(see Section 4.3), set the default output required between –100
Controller
Dual
and 100% in 0.1% increments.
Notes.
• A setting of 1 to 100% represents a direct output.
• A setting of –1 to –100% represents a reverse output.
• A setting of 0% represents no output i.e. both direct and reverse outputs are switched
off.
Return to main menu.
CONFIG. CONTROL
Optional analog output board fitted – see User Guide, Section 5.6.
CONFIG. CLOCK
Optional analog output board not fitted – see User Guide, Section 5.7.
CONFIG. SECURITY
(see Section 2.2) is set automatically
Control Mode
is set to
irrespective of its setting prior to the power
Auto
is set to
irrespective of its setting prior to the power
Manual
(see Section 2.2) is set to the level set in the
and
are set to the same state as that set prior
Control Output
PROGRAMMING
Default
19