Section 1
Specifications
Range
Accuracy
Resolution (displayed)
Repeatability
Response time
Sample flow required
Sensor storage
temperature
Operating temperature
Sample temperature range
Operating humidity
Power requirements
Sample inlet fitting
Overflow drain fitting
Body drain fitting
AIr purge fitting
Signal average (filter) time
Sensor dimensions
Sensor cable length
Sensor cable rating
Mounting options
Shipping weight
Calibration method
Verification (dry) method
Recommended cleaning
intervals
Languages
Installation environment
Primary compliance
method
Specifications are subject to change without notice
0.01–9999.9 nephelometric turbidity units (NTU)
± 5% of reading or ± 0.1 NTU (whichever is greater) from 0.01 to 2000 NTU;
± 10% of reading from 2000 to 9999 NTU
0.01 NTU up to 999.99 NTU; 0.1 NTU from 1000.0 to 9999.9 NTU
Better than ± 1.0% of reading or ± 0.04 NTU, whichever is greater for each range.
Initial response in 45 seconds
1.0 to 2.0 L/min (0.3 to 0.5 gal/min) (15 to 30 gal/hr)
–20 to 80 °C (–4 to 140 °F); 95% relative humidity, non-condensing.
sc100: 0 to 50 °C (32 to 122 °F) for one SS7 sc on an sc100; 0 to 40 °C (32 to 104 ° F) for one
SS7 sc and another smart sensor that consumes less than 5 watts on a single sc100
(provided by sc100). See
Figure 1 on page
sc1000: –20 to 55 °C (–4 to 131 °F); 95% relative humidity, non-condensing
0 to 50 °C (32–122 °F);
HST model—0 to 70 °C, intermittent 70 to 80 °C. (An approved heat exchanger is available to
reduce sample temperature.)
5 to 95% non-condensing
sc100: 12 VDC ±5%, 20 watts maximum
sc1000: 100–240 V ± 10 VAC, 50/60 Hz from 75 VA standard to a maximum of 2000 VA (only
when sc analyzers are connected), Category II or 24 VDC (18–30 VDC), maximum 75 W
¾-in. NPT female
1-in. NPT female
¾-in. NPT female
¼-in. quick-connect compression fitting; 0–50SCFH airflow of clean instrument air
No averaging, 6, 30, 60 and 90 seconds, user selectable. Default is 30 seconds.
64.2 x 67.5 x 19.0 cm (25.3 x 26.6 x 7.5 in.)
2 m (6.6 ft); Optional 7.62 m (25 ft) extension cable. Maximum cable length is 9.62 m (31.6 ft).
Cable: 105 °C, 300 V, PVC jacket
Wires: 22 AWG, PVC jacket
Wall
SS7 sc—15.8 kg (34.8 lb);
SS7 sc-HST—18 kg (39.6 lb)
Formazin – user-prepared primary or wet calibration of the instrument
Standardization plates with approximate values of 100 or 1000 NTU. Unique value is assigned
when dry verification is done immediately after calibration and is used with pass/fail criteria for
subsequent verifications.
Mandatory before calibration
Optional before verification
Mandatory upon verification failure
English (default), German, French, Spanish, Italian, Swedish, Polish, Korean,
Chinese, Japanese
Indoor
USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B
6.
5