Section 1 Specifications

Specifications are subject to change without notice.
General specifications
Specification
Dimensions (W x D x H)
Enclosure rating
Weight
Installation environment
Mount
Pollution degree/installation
category
Protection class
Altitude
Light source
Detector
Optical path length
Time delay
Trip point
Reagents
Reagent shelf life
Power requirements
Electrical connections
Hard water alarm relay
Contact rating
Operating temperature
Storage temperature
Operating humidity
Sampling rate
Details
42 x 31.5 x 18 cm (16.5 x 12.5 x 7 in.)
IP62
11.3 kg (25 lbs)
Indoor
Wall
2/II
I
Maximum 2000 m (6562 ft)
Class 1 LED (light emitting diode) with peak wavelength of
610 nm. Estimated 50,000 hours minimum life; hard water
and soft water indicator.
Silicon photo detector
8.89 mm (0.35 in.)
One reading above the trip point is necessary to trigger a
soft-to-hard water alarm transition. One reading below the
trip point is necessary to trigger a hard-to-soft water alarm
transition.
0.3, 1, 2, 5, 10, 20, 50 and 100 mg/L. Refer to
specifications
on page 4.
Buffer and indicator solution. 500 mL of each reagent every
2 months.
1 year
115/230 VAC, 70 VA, 50/60 Hz, 1.25 A Fuse
Two 3-wire barrier terminal blocks. Wire range: 18–12 AWG
2
(0.75–1.0 mm
)
SPDT relay, set to on when the hard water indicator is on.
The alarm can be disabled.
5 A resistive at 100–240 VAC
5 to 40 °C (32 to 104 °F)
-40 to 60 °C (-40 to 140 °F)
Relative humidity: 5–95% at different temperatures, non-
condensing
New sample: every 1.9 minutes ±5% at 60 Hz; 2.3 minutes
±5% at 50 Hz