KERN CGB Operating Instructions Manual page 48

Counting balances
Table of Contents

Advertisement

Available languages

Available languages

Selector
– Turn this selector to set the reference
sample quantity.
– Each time you change the selector
setting, the new reference sample
quantity will be briefly displayed.
Reference Sample Updating
Automatic reference sample updating
optimizes the counting accuracy.
You can activate or de-activate this
function in the menu.
The abbreviation opt for "optimizing"
briefly appears in the "PIECE WEIGHT"
display to indicate that reference sample
updating has been completed if
– the criterion for the stability parameter
selected in the menu has been met
– the current piece count is less than twice
the original piece count
– the current piece count is less than
1,000
– the internally calculated piece count
(such as 17.24 pcs) differs by less than
± 0.3 pcs from the whole number
(17 pcs in this example)
PK Elektronik Vertriebs GmbH, E-Mail: info@pkelektronik.com, Internet: www.pkelektronik.com
48
Factory Settings of the Parameters
Program selection:
Counting without weight unit toggling
(2. 1. 4)
Resolution:
Standard: according to the displayed
accuracy
(3. 4. 1)
Storage parameter:
At stability (3. 5. 1)
Reference sample updating:
Automatic (3. 6. 2)
Autotare when [START] is pressed:
On (3. 7. 2)
Additional Functions
In addition to the basic functions
(power off, zeroing, taring and printing),
you can also access the following
functions from this application:
Cancel initialization:
[DELETE] key
Show weight: [UNIT] key
Calibrate/adjust balance:
press [TARE] > 2 sec.