Technical Specifications

Operating temperature:
Storage temperature:
Weight:
Dimensions:
Measuring procedure:
Measurement range:
Cuff pressure display:
Power source:
10 - 40 °C / 50 - 104 °F
15 - 85 % relative maximum humidity
-20 - +50 °C / -4 - +122 °F
15 - 90 % relative maximum humidity
260g (including batteries)
115 x 80 x 35 mm
oscillometric, corresponding to
Korotkoff
30 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within
±3 mmHg
Pulse accuracy: ±5 % of the readout
value
4X1.5 V Batteries; size AAA
Reference to Standards:
EU Directives 93/42/EEC
NIBP requirements:
EN 1060-1 / -3/ -4,
EN 60601-2-30,
ANSI /AAMI SP10
Microlife reserves the right to alter technical
specifications without prior written notice.
39