D.1.1.7 Lamp Timer Function - Fujitsu M3096GX Instruction Manual

Image scanner
Table of Contents

Advertisement

D.1.1.7
Lamp timer function
f
Functions
-
This function specifies how long the lamp will stay on.
-
This must be specified with Mode select (Page code=Xπ3Dπ).
Format of Page Code Xπ3Dπ
Byte 0
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Byte 6
Byte 7
f
Parameter
Xπ00π: The default value is 60 seconds for this scanner.
Xπ01π to XπFFπ: Specifiable between 1 to 255 seconds (XπFFπ)
D.1.1.8
Added sense code
Following sense code are added.
The blown lamp fuse;
™Sense key=Xπ04π, Add.sense code=Xπ80π, Add.sense qualifier=Xπ03π∫.
Interlock switch is opened;
™Sense key=Xπ02π, Add.sense code=Xπ80π, Add.sense qualifier=Xπ01π∫.
D $ 10
Reserved
Parameter length = Xπ06π
Parameter
Reserved
Byte 2
Page code = Xπ3Dπ