13-4 CAMERALINK INTERFACE SPECIFICATIONS

CameraLink interface is a standard of a digital interface for industrial application
standardized by digital camera manufactures and frame grabber board manufactures. 28
bits digital data (TX0 to TX27) can be transferred as being changed into 5 signals (X0, X1,
X2, X3 and XCLK) by parallel serial transformation. This camera is based on CameraLink
interface and Base Configuration 14 bit digital camera standard.
(1) CameraLink connector pin assignments (MDR-26)
Camera connector
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
Note
Note
Frame grabber connector
1
25
24
23
22
21
20
19
18
17
16
15
13
14
12
11
10
9
8
7
6
5
4
3
2
26
CC1 to CC4 are options. It is possible to use an external trigger.
So when it is needed, please feel free to contact our sales
group.
SerTC and SerTFG are available when the second of DIPSW
on the back panel of the camera is off.
C9100-02 Ver.1.7
Channel Link signal
Inner Shield
X0-
X1-
X2-
Xclk-
X3-
SerTC+
SerTFG-
CC1-
CC2+
CC3-
CC4+
Inner Shield
Inner Shield
X0+
X1+
X2+
Xclk+
X3+
SerTC-
SerTFG+
CC1+
CC2-
CC3+
CC4-
Inner Shield