(3) Synchronous Readout trigger mode (pulse polarity is negative)
Readout starts when Trigger In pulse is detected. The exposure time is defined as
the time span of Trigger In pulses. The first pulse is ignored when this trigger mode
is selected.
Trigger In
Exposure
Readout

13-5-3 INTEG OUT TIMINGS

The integration-timing signal comes out. The signal level is C-MOS 3.3 V and it is available
only in internal mode.
Integ Out
Exposure
24
Trigger In cycle
Over 1 µs
Within 925 µs
Exposure
Readout
10 µs
Within 64 µs
Exposure time
Exposure
+3.3 V
0 V