Adjusting The Gap Sensor Level Manually - SATO S84 Operator's Manual

Table of Contents

Advertisement

5.5.3

Adjusting the Gap Sensor Level Manually

The adjustment procedure for the Gap sensor level is as follows:
1
Go to the SENSOR LEVEL setting screen
of SERVICE MODE.
Perform steps 1 through 4 of
Adjusting the Media Sensor
2
Select MANUAL using the
and then press the
SENSOR SELECT shows on the screen.
3
Press the
buttons to select GAP and
/
then press the
The light emission level adjustment of the Gap
sensor is shown.
4
On the SENSOR LEVEL GAP(E) screen,
press the
buttons to set ADJUST
/
LEVEL to 2 and then press the
button.
The adjustment range of ADJUST LEVEL is from 0
to 3.
We recommend using the initial value which is 2.
The light reception level adjustment of the Gap
sensor is shown.
First, adjust the "Low" reception level (voltage) of the Gap sensor.
5
Open the top cover and unlock the media
feed and media sensor assembly .
6
Physically place the liner  without the
label resting over the Gap sensor.
7
Press the media feed and media sensor
assembly  down to lock it.
To get the correct adjustment result, close the
media sensor assembly before performing the
adjustment.