Fig. 38
@
Fig. 39
ƒ
|
Fig. 40
9
Attaching the Eyepiece
# When using the U-BI30-2 binocular tube, eyepieces with built-in mi-
crometer disk cannot be attached.
# When using a finder eyepiece or an eyepiece with focusing mirror,
@
attach it to the right-hand eyepiece sleeve.
When doing so, make sure that the eyepiece positioning pin @ fits
²
into the notch ² on the eyepiece sleeve.
# The super-widefield trinocular observation tube is equipped with a
positioning notch on both eyepiece sleeves. Make sure that the
positioning pins on both eyepieces fits into the respective notches.
10
Attaching the Power Cord
The power cord is vulnerable when bent or twisted. Never subject it
to excessive force.
Make sure that the main switch @ is set to "
³
necting the power cord.
Always use the power cord provided by Olympus. IF no power cord
is provided with the microscope, please select the proper power
cord by referring to section "PROPER SELECTION OF THE POWER
²
SUPPLY CORD" at the end of this instruction manual.
1. Connect the power cord plug ² to the AC receptacle ³.
The power cord should be connected to a grounded, 3-conductor
power outlet. If the power outlet is not grounded properly, Olympus
can no longer warrant the electrical safety performance of the equip-
ment.
2. Plug the power cord plug | into the wall outlet ƒ.
Should the power cord come in contact with the hot lamp housing,
the power cord could melt and cause electric shock. Be sure to
route the power cord away from the lamp housing.
BX51/BX52
(Fig. 39)
(Figs. 40 & 41)
" (OFF) before con-
29