Auxiliary Interface; Inputs - Datalogic DX6400 Reference Manual

Hide thumbs Also See for DX6400:
Table of Contents

Advertisement

INSTALLATION

Auxiliary Interface

The auxiliary serial interface is equipped with RS232 full-duplex interface connections. The
interface type is exclusive and is selectable through the Genius™ configuration program.
The following pins of the 25-pin and 26-pin connector are used for RS232 full-duplex
interface connection:

Inputs

The inputs of the reader are on the 25-pin and 26-pin connector (Figure 4, 1 and Figure 5, 1)
of the DX6400/DX6500.
These inputs are called EXT_TRIG, IN2, IN3 and IN4.
Pin
Name
18
EXT_TRIG A External trigger (polarity insensitive)
19
EXT_TRIG B External trigger (polarity insensitive)
6
IN2A
10
IN2B
14
IN3A
15
IN4A
24
IN_REF
IN2 is normally used for the Encoder input. In PackTrack™ mode, it detects the conveyor
speed.
EXT_TRIG is the main presence sensor. When active, this input tells the scanner to scan for
a code and that decoding can take place. The yellow LED (Figure 3,3) indicates the
EXT_TRIG is active.
IN3 and IN4 can be used as the stop signal for the reading phase.
All inputs are optocoupled, polarity insensitive, and driven by a constant current generator;
the command signal is filtered through an anti-disturbance circuit which generates a delay
Pin
Name
20
RXAUX
21
TXAUX
23
SGND AUX
DX6X00
20
21
23
1
Shield
Figure 16 - RS232 Auxiliary Interface Connections
Function
Input signal 2 (polarity insensitive)
Input signal 2 (polarity insensitive)
Input signal 3 (polarity insensitive)
Input signal 4 (polarity insensitive)
Common reference of IN3 and IN4 (polarity insensitive)
Function
Receive data
Transmit data
Auxiliary signal ground
USER INTERFACE
RXAUX
TXD
TXAUX
RXD
GNDAUX
GND
Earth
Ground
2