Procedure
• Ensure that the preload for which the calibration zero point is to be moved is in place.
• Start the function. You will see a prompt asking you if you really want to move the calibration zero point. The calibration
zero point is moved when you confirm the prompt.
If the display signals an underload or overload when you quit this menu, switch the scale off and then on again.
7.13
Defining MinWeigh value (SCALE–> Minweigh)
The MinWeigh function is to ensure that the weighing results lie within the defined tolerances corresponding to the
specifications of the quality-assurance system used.
If the value for MinWeigh = 0, the menu for the supervisor is fully hidden.
The value can be entered with the numerical keypad for KMT-TM terminals (ITT). Press the
and KMB-/KMS-TM terminals (ITB/ITS). The first digit starts flashing and can now be changed with the
Confirm new digit with the
7.14

Settings in the "TERMINAL" menu block

Most of the TERMINAL menu block is also available to the supervisor and is therefore not described here (see User Manual).
However, this menu block includes a submenu that is restricted to the service technician and is described below.
Defining a timeout ( TERMINAL –> Timeout )
7.14.1
In this menu you can limit the period of time for which the scale is used. This is particularly useful for leased scales, since it
prevents the scale for being used for longer than the agreed period. This menu item is only available for KMT-TM terminals
(ITT). The following settings are available:
"Not active"
"Leasing"
7.15

Setting of country version

After replacing the analog or digital board of the scale it is absolutely necessary to check the country specific settings and to
adjust them if needed. The country specific settings affect the Technician Password, keyboard map, etc. Technician Password
EU =
For this procedure please connect the scale to an PC with an appropriated cable (PC – Scale, RS232, 9-pin straight cable
Sub D m/f) and open a terminal program on your PC (e.g. HyperTerminal). Verify that the scale is in "Dialog" Mode and the
settings of both interface of the scale and PC correspond.
Depending on your location you may need to send the SICS command "DSK_US" (for USA and Canada) or "DSK_EU" (for all
other countries) to the scale. The scale will send back "DSK_A". This means that the command had been executed
successfully. If this is not the case, please re-check the connection between scale and PC as well as the interface settings and
repeat this procedure again. To inquire the current status of the "DSK", send the SICS command "DSK". Note that each SICS
command must be terminated with CRLF (carriage return, linefeed).
60
key.
There is no restriction on the period of use of the scale (factory setting).
The duration of use is limited to the number of days entered. Enter the number of days (1 - 999)
during which the scale can be used in the "Days" menu item. The duration of use is valid as of the
next day. If the set duration of use has passed, the scale still starts up, but no longer performs
measurements and instead displays the message that the duration of use has passed.
, US =
.
key for compact scales FTx
and
keys.
FTx/ITx-SH-e-0720