DigiCal DCS 2
5.

Technical data

Type:
Probe Ports 1 / 2
Measuring range
Accuracy
Pt 100 temperature probe with
manufacturer calibration
Resolution
Outputs
Connectors
Allowable operating temp.
Display
Housing
Dimensions
Weight
Power supply
DCS 2 Ref. No.
YAME0017/ 09.10.03
DigiCal DCS 2
-200...–100°C: 0.05°C
-100...+150°C: 0.03°C
+150...+200°C: 0.05°C
-200 bis +200°C with 0.01°C resolution
+200 bis +400°C with 0.1°C resolution
RS232 Interface (serial Printer can be connected)
2 analogue outputs 0-1 V (11-bit scalable)
200 x 80 x 200 mm (BxHxL)
Power Supply 230V; 50/60Hz or optional 90-260V; 50/60Hz
LMD 018 (230V)
LMD 919 (90-260V)
Technical data
Pt100
-200...+450°C
>200°C: 0.1%
DIN 8-pole
0°C ... +40°C
2-line LCD
plastic (ABS)
950 g