Calibration Function - B @ K - Lauda DCS2 Operating Instructions Manual

Table of Contents

Advertisement

DigiCal DCS 2
2.4.3 Calibration function -
Despite high quality manufacturing techniques, each probe is slightly different from
specified standards. To eliminate inaccuracies caused by exchanging or ageing of probes,
the instrument offer easy calibration functions which guarantee that the system accuracy
is always as good as if the instrument was specifically calibrated to the individual probes in
our laboratory.
The instruments offer three calibration options:
= calibration
B @ K
Push [ENTER/MENUE] to calibrate the instrument with sensor. Use the up and down keys [
to select
B @ K
indicates the selected channel. For changing the channel use the up and down keys [
[ENTER/MENUE] to confirm..
Use the up and down keys [
Display
B @ K
1
B @ K
B @ i
2
1
B @ i
1
B @ i
m O 0
m O 1
[
m E E
YAME0017/ 09.10.03
B @ K
[
]
: LAUDA-factory calibration, encoded calibration with optimum accuracy.
m O 0
Probe and electronics are being calibrated at the same time. Therefore the
accuracy reaches 0,03°C for temperature range –100°C to +150°C.
This calibration is stored as
Please note that the accuracy will be 0,05°C in case another probe is connected
that was calibrated without the electronics.
(This probe data can be read as a 2x4 digits code on the handle of the LAUDA
DigiCal probes and results from the 2-point-comparison method
[
]
: Calibration by physical standard references with the best possible system
m O 1
accuracy for a user's calibration (calibration with comparative standards: 1-point,
2-point or 3-point-comparison possible).
[
]
: Standard characteristic curve with a minimum system accuracy
m E E
(e.g. Pt100-resistance according DIN IEC 60751).
. Push again [ENTER/MENUE]. On the left corner the display appears a small 1, which
] to select the requested calibration option.
Keys
ENTER/MENUE
up and down keys [
ENTER/MENUE
up and down keys [
ENTER/MENUE
ESC
calibration code.
m O 0
Function
select the channel 1 or 2
]
select the calibration mode
]
To keep the amended value
To escape without retention
Handling / Introduction
.
m O 0
]
]. Push