Strip Chart Commands - Stanford Research Systems SR865 Operation Manual

2 mhz dsp lock-in amplifier
Table of Contents

Advertisement

120
Programming

Strip Chart Commands

GSPD(?) { i }
The
GSPD i
Example
GSPD 6
GSPD?
GSCL(?) [ DAT1 | DAT2 | DAT3 | DAT4 | j ] { , x }
The
GSCL j, x
be set to the nearest allowed scale in a 1-2-5 sequence.
The value of j (0–3) corresponds to the
(orange) data channels.
DAT4
Example
GSCL DAT2, 0.1
GSCL 1, 0.09
GSCL 1, 0.12
GSCL? 1
GSCL? DAT2
GOFF(?) [ DAT1 | DAT2 | DAT3 | DAT4 | j ] { , x }
The
GOFF j, x
The value of j (0–3) corresponds to the
(orange) data channels.
DAT4
Example
GOFF DAT3, 0.1
GOFF 2, 0.1
GOFF? 2
GOFF? DAT3
SR865 DSP Lock-in Amplifier
command sets the horizontal time/div according to the table below.
i
time/div
0
0.5 s
1
1 s
2
2 s
3
5 s
4
10 s
5
30 s
6
1 min
7
2 min
8
5 min
Set the strip chart to 1 min/div horizontal scale
Returns the horizontal scale i
command sets the vertical scale of data channel j to x/div. The scale will
Set the vertical scale for data channel 2 (blue) to 0.1/div
Returns the vertical scale for data channel 2 (blue)
command sets the vertical offset of data channel j to x.
Set the vertical offset for data channel 3 (yellow) to 0.1
Returns the vertical offset for data channel 3 (yellow)
i
time/div
9
10 min
10
30 min
11
1 hour
12
2 hour
13
6 hour
14
12 hour
15
1 day
16
2 day
(green),
(blue),
DAT1
DAT2
(green),
(blue),
DAT1
DAT2
Chapter 4
(yellow) and
DAT3
(yellow) and
DAT3

Hide quick links: