System Commands - Stanford Research Systems SR865 Operation Manual

2 mhz dsp lock-in amplifier
Table of Contents

Advertisement

Chapter 4

System Commands

TIME(?) [ SEConds | MINutes | HOUrs | j ] { , i }
The
TIME j, i
(j=2) to i.
Example
TIME MIN, 30
TIME 1, 30
TIME? MIN
DATE(?) [ DAY | MONth | YEAr | j ] { , i }
The
DATE j, i
The value of i is limited to 0–99.
Example
DATE YEAR, 21
DATE 2, 21
DATE? YEAR
TBMODE(?) { AUTO | INternal | i }
The
TBMODE i
internal (i=1).
Example
TBMODE AUTO
TBMODE 0
TBMODE?
TBSTAT?
The
TBSTAT?
internal (1).
Example
TBSTAT?
BLAZEX(?) { BLazex | BIsync | UNIsync | i }
The
BLAZEX i
(i=1) or unipolar (positive) sync (1=2).
Example
BLAZEX UNI
BLAZEX 2
BLAZEX?
command sets the clock value for seconds (j=0), minutes (j=1) or hours
Set the system clock minutes to 30
Returns the system clock minutes
command sets the clock value for day (j=0), month (j=1) or year (j=2) to i.
Set system clock year to (20)21
Returns the system clock year (last 2 digits)
command sets the external 10 MHz timebase mode to auto (i=0) or
Set the timebase mode to auto
Returns the timebase mode i
query returns the current 10 MHz timebase source, either external (0) or
Returns the current 10 MHz timebase
command sets the rear panel BlazeX output to blazex (i=0), bipolar sync
Set the BlazeX output to unipolar sync
Returns the BlazeX output selection i
Programming
SR865 DSP Lock-in Amplifier
143

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents