PROGRAMMING A REPLACEMENT CONTROL
BOARD
Indigo™ replacement control boards require the Model
number to be entered to activate the appropriate look
up tables for operation and diagnostic. This can be
done two different ways, USB Setup or Manual Setup.
USB Setup - Applicable when the control board is
operational and has a mechanical issue such as a
sticking relay. The asset data is transferred to the
replacement control board from the faulty control
board. Refer to page 98 "Exporting Data To A Flash
Drive" before installing the replacement board.
Manual Setup - Applicable when the control board is
non-operational or data from the faulty board is
suspect.
Install replacement control board and reapply power.
G series (blue) control boards will go directly to the
Replacement Control Board screen and pause for
input - Select either "Manual" or "USB" setup
option.
E series (beige) control boards require navigation
to the Replacement Control Board (Repl Cntl Bd)
screen.
1. Press the Menu button.
2. Press the Down arrow until Service is highlighted,
then press the Checkmark.
3. Scroll down to Replacement Control Board and
press the Checkmark.
4. Select "Manual Setup" and press the Checkmark.
Verify the date correctly auto populates.
Yes - Go to step 5
No - Press the right arrow to activate the cursor,
then use up/down arrows to enter the correct date.
Part Number STH025 8/12
165