Initial Test Procedures; Testing The Motor - Invacare Pronto M41 Series Service Manual

Hide thumbs Also See for Pronto M41 Series:
Table of Contents

Advertisement

®
Invacare
- PRONTO M41
7

Initial test procedures

7.1

Testing the motor

On the 4‐pin motor connector (2), locate the two contacts in the RED and BLACK housings.
Set the digital multimeter to measure resistance (ohms).
Measure the resistance between the two motor contacts.
Note
A normal reading is between 0.5 and 5 ohms. A reading of 0 ohms or in excess of 15 ohms
indicates a problem. High readings are generally caused by bad connections and/or damaged
brushes. Contact Invacare.
SERVICE MANUAL
31