SERVICE MANUAL
·
Carefully note the location of the cable and the connection locations of the various plugs. Either
mark each plug and socket, or take a photograph with a digital camera.
·
In the illustration above, you see the power module (3) and the positions of the optional lighting
circuit board (2), of the optional actuator module (1) and of the optional G-Trac sensor (4).
·
Pull the plug (1) out of the power module.
·
Use a 4 mm Allen wrench to loosen both of the hexagon socket head screws (2).
·
Remove the washers and the black intermediate pieces.
·
Remove the power module by pulling it backwards from the guides.
·
Replace the power module.
Refitting the power module:
·
Refit the parts in reverse order.
·
Plug all the plugs into their old positions.
·
Select the drive program as described in chapter 8.2.6.
·
If a new software version is available, update the drive program as described in chapter 8.3.
44
®