6.1

Drive fault diagnosis

PROBLEM
Mobility aid
will not
start
Mobility aid
judders in
drive mode
Mobility aid
pulls to left
or right
Error
message
does not go
out
4
MAX
OTHER
POSSIBLE
SYMPTOMS
CAUSE
The remote
Drive motors
status display
disengaged
illuminates
normally and
displays blink
code 5/6.
Remote status
Batteries
display does not
defective
illuminate
Completely
discharge
battery
Power supply to
remote
interrupted
Remote
defective
Remote status
Various causes
display blinking
None
Batteries
defective
(unstable
voltage)
Drive motor(s)
defective
None
Drive motors
running
asymmetrically
Tyre visibly
Not enough air in
dented
the tyre
None
Bad connections Check all
Motor brake
defective
SOLUTION
Documentation
See operating manual
Engage drive
motors
Replace batteries
See chapter 8.4.2.
Pre-charge
See instruction manual for
batteries
charger
Check master fuse See chapter 8.5.
Check cables
See chapter 8.6.
between the
modules for loose
connections and
damage
Replace remote
See chapter 8.2.1.
Assess error code See chapter 6.2.
Replace batteries
See chapter 8.4.2.
Replace motor(s)
See chapter 8.11.1.
Replace carbon
See chapter 8.11.2.
brushes
Change
See chapter 8.3.
programming to
synchronise
motors.
Check air
See chapter 8.10.2.
pressure, replace
inner tube and/or
valve if necessary.
See chapter 8.6.
connecting cables.
Measure internal
See chapter 8.11.1
resistance of
brakes, replace
motor if defective.
SERVICE MANUAL