Manufacture's Declaration
Electromagnetic Immunity (IEC60601-1-2)
Immunity
Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on
power supply
IEC 61000-4-11
Power frequency
(50/ 60 Hz)
magnetic field
IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.
34
IEC60601-1-2
Test Level
±2,4,6 kV contact
±2,4,6 kV contact
±2,4,8 kV air
±2,4,8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV normal
±1 kV normal
mode
mode
±2 kV common
±2 kV common
mode
mode
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 0.5 cycle
for 0.5 cycle
40 % UT
40 % UT
(60 % dip in UT)
(60 % dip in UT)
for 5 cycles
for 5 cycles
70 % UT
70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles
for 25 cycles
<5 % UT
<5 % UT
(95 % dip in UT)
(95 % dip in UT)
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Compliance
Electromagnetic Environment
Level
Floor should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
If the user of the HBP-9020 requires
continued operation during power mains
interruption, it is recommended that the
HBP-9020 be powered from an
uninterruptible power supply.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Guidance