Omron HBP-1100 Instruction Manual

Omron HBP-1100 Instruction Manual

Professional blood pressure monitor
Hide thumbs Also See for HBP-1100:

Advertisement

Professional Blood Pressure Monitor
HBP-1100
Instruction Manual
Thank you for purchasing this OMRON Professional Blood
Pressure Monitor.
Please completely read this Instruction Manual before using the
monitor for the first time.
Read this manual to ensure the safe and accurate use of the
monitor.
XXXXXXX-X A

Advertisement

Table of Contents
loading

Summary of Contents for Omron HBP-1100

  • Page 1 Professional Blood Pressure Monitor HBP-1100 Instruction Manual Thank you for purchasing this OMRON Professional Blood Pressure Monitor. Please completely read this Instruction Manual before using the monitor for the first time. Read this manual to ensure the safe and accurate use of the monitor.
  • Page 2 OMRON or the dealer specified by OMRON 2. The problem or damage of OMRON product caused by the product of other manufacturer not delivered by OMRON 3. The problem and damage caused by the maintenance and/or repair using the repair parts not specified by OMRON 4.
  • Page 3: Table Of Contents

    Table of Contents Introduction Intended Use ................2 Features of the Product ..............2 Notes on Safety ................3 Using the Unit Components of the Product............10 Features / Functions of Unit ............11 Installing the Batteries ..............13 Connecting the AC Adapter............13 Cuff Selection and Connection .............14 Applying the Cuff to the Patient ............15 Zero indicator function ..............17 Non-Invasive Blood Pressure (NIBP) Measurement...
  • Page 4: Intended Use

    Features of the Product The blood pressure accuracy of the HBP-1100 is clinically proven. Easy to use, the HBP-1100 is intended for use by medical professionals. ■ Zero indicator function (page 14): Before each measurement, this device indicates that "zero calibration"...
  • Page 5: Notes On Safety

    • Do not use the cuff or AC adapter to lift the unit, it can also cause the unit to malfunction. • If the unit has broken down, contact your local OMRON representative. • Do not use in combination with a hyperbaric oxygen therapy device, or in an environment where combustible gas may be generated.
  • Page 6: Using The Unit

    • For the AC adapter connected to the unit, supplies, and optional devices, use only the standard accessories or OMRON-specified products. • Do not use in a location with moisture, or a location where water may splash on the unit.
  • Page 7: Using The Unit

    Caution • Before using the unit, verify that none of the following apply to the patient: - Poor peripheral circulation, noticeably low blood pressure, or low body temperature (there will be low blood flow to the measurement position) - The patient uses an artificial heart and lung (there will be no pulse) - An SpO sensor and the cuff are attached to the same arm - The patient has an aneurysm...
  • Page 8 Cleaning Warning • When cleaning the unit, turn off the power and disconnect the AC adapter from the unit. • After cleaning the unit, make sure it is completely dry before connecting to a power outlet. • Do not spray, pour, or spill liquids into or onto the unit, accessories, connectors, buttons, or openings in the housing.
  • Page 9 Dry cell battery Warning • If battery fluid comes in contact with the eye, immediately flush with copious amounts of water. Do not rub. Seek medical attention immediately. • Do not throw into flame, disassemble, or heat. • Always disconnect the AC adapter from the unit before removing or installing a battery.
  • Page 10 • Do not use the cuff if it is damaged or has holes. • Only an OMRON GS CUFF can be used with this device. The use of any other cuff may result in incorrect measurement.
  • Page 11 Note: Setup • Read and understand the manual for each optional accessory. This manual does not contain cautionary information for optional accessory. • Exercise caution with the cables and arrange so that the patient does not become entangled or bound. Before use / during use •...
  • Page 12: Using The Unit Components Of The Product

    OMRON representative. Main unit Standard Medical Accessories AC adapter GS CUFF M Instruction Manual Caution • Only an OMRON GS CUFF can be used with this device. The use of any other cuff may result in incorrect measurement.
  • Page 13: Features / Functions Of Unit

    Features / Functions of Unit Front and bottom of unit [Power ON/OFF] Turns the power ON/OFF. button Press to enter auscultation mode. scultatio ] button (page 20). Press to start blood pressure measurement. [START/STOP] While the cuff is inflating, hold button down to continuously inflate (page 18).
  • Page 14 LCD Display Displays systolic blood pressure. Displays diastolic blood pressure. Pulse Displays the pulse rate. Lights in the measurement result display if Irregular pulse the pulse wave interval was irregular or wave icon there was body movement during measurement. Auscultation icon Lights when "Auscultation Mode" is ON. Pulse Flashes in synchronization with the pulse synchronization...
  • Page 15: Installing The Batteries

    Installing the Batteries Warning • If battery fluid comes in contact with the eye, immediately flush with copious amounts of water. Do not rub. Seek medical attention immediately. • Do not throw into flame, disassemble, or heat. • Do not attempt to disassemble or modify the battery. •...
  • Page 16: Cuff Selection And Connection

    If too small a cuff is used, the measured blood pressure value tends to be higher. • Only an OMRON GS CUFF can be used with this device. The use of any other cuff may result in incorrect measurement.
  • Page 17: Applying The Cuff To The Patient

    Applying the Cuff to the Patient The device can be used on either the right or left arm. Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a rolled up sleeve will cause inaccurate blood pressure measurements.
  • Page 18 Caution • Make sure the cuff is wrapped in the correct arm position and is at the same height as the heart. NOTE: • If measurement is difficult due to arrhythmia, use a different blood pressure measurement method. • If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where the cuff is to be wrapped, follow the instructions of the doctor.
  • Page 19: Zero Indicator Function

    Zero indicator function Before each measurement, this device indicates that “zero calibration” was successful. ■ When the power is turned on, “zero calibration” starts after the entire indicator blinks. When completed, appears. ■ When the power is already on, “zero calibration” takes place from the ready screen (which shows “0”).
  • Page 20: Non-Invasive Blood Pressure (Nibp) Measurement Measurement In Normal Mode

    Non-Invasive Blood Pressure (NIBP) Measurement Measurement in normal mode 1. Press the [START/STOP] button. Blood pressure measurement is performed once. 2. The measurement results are displayed. If a measurement value is outside the corresponding range below, the value will flash. SYS: 59 mmHg or less, or 251 mmHg or higher.
  • Page 21 Caution • The accuracy of a flashing measurement value that is outside the measurement range is not guaranteed. Always check the patient's condition before deciding what steps to take. NOTE: • If inflation is insufficient, inflation may restart automatically while measurement is in progress.
  • Page 22: Measurement In Auscultation Mode

    Measurement in auscultation mode In auscultation mode, this device does not measure blood pressure. Measurement should be performed by a health care professional using a stethoscope. The healthcare professional uses a stethoscope to determine SYS and DIA by means of the auscultation method. 1.
  • Page 23: Stopping The Measurement

    NOTE: • The body movement detection function is disabled while "Auscultation Mode" is in use. • In auscultation mode, the pulse rate is not measured and does not appear. Stopping the Measurement To stop measurement while measurement is in progress, press [START/ STOP] button.
  • Page 24: Non-Invasive Pressure Measurement Principles

    Non-Invasive Pressure Measurement Principles Oscillometric method The beat in the pulsation generated by the contraction of the heart is captured as the pressure inside the cuff to measure the blood pressure. If the cuff wrapped around the upper arm is pressurized sufficiently, the blood flow stops, but the beat of the pulsation is present and the pressure inside the cuff receives this and oscillates.
  • Page 25: Maintenance

    Maintenance Maintenance Inspection and Safety Management The HBP-1100 must be maintained to ensure functionality and to secure the safety of patients and operators. Daily checks and maintenance should be performed by the operator. In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct periodic inspections.
  • Page 26: Check Before Use

    • Press each button and check that it works. ■ Non-invasive blood pressure (NIBP) • Make sure that a suitable OMRON GS CUFF is attached (one that fits the circumference of the patient's arm). • The cuff tube is firmly connected.
  • Page 27: Checking Pressure Accuracy

    Checking pressure accuracy You can check the pressure accuracy of the device. [Power ON/OFF] [START/STOP] button button [Auscultation] button 1. Make sure the power is on. 2. Hold down the [Auscultation] button for 3 seconds. "Zero calibration" is performed. When calibration is finished, the pressure accuracy verification screen appears.
  • Page 28: Troubleshooting

    AC adapter, remove the batteries, and contact your local OMRON representative. The unit display does not operate Cause / solution Stop using the unit and contact your local OMRON representative. The unit becomes hot Cause Solution...
  • Page 29 Measurement was not possible Cause / solution Check the patient by palpation or other method. After checking the patient, check the error code and see "List of Error Codes" (page 29) for Non Invasive Blood Pressure (NIBP) measurement. Abnormal measurement value Cause / solution The causes below are possible.
  • Page 30 The measurement value is questionable Cause Solution Deflates quickly Check for a loose cuff connection. Measure circumference or patient's Incorrect cuff size used. arm and ensure correct sized cuff is used. Ensure cuff is applied to a bare arm, Cuff wrapped over thick clothing. or very thin clothing.
  • Page 31: List Of Error Codes

    Points to check code The cuff tube is not connected Firmly connect the cuff tube. Replace with an OMRON GS Air is leaking from the cuff. CUFF that does not leak. Did not inflate properly because Have the patient not move the...
  • Page 32 If this occurs during measurement, repeat measurement. Over-inflation occurs If this occurs when not performing measurement, contact your local OMRON representative. Replace with new batteries. The batteries are depleted. (page 13) Replace the batteries with new batteries. If the error continues,...
  • Page 33: Disposal

    Disposal As there is a risk of environmental pollution, follow your applicable national and local legal regulations regarding disposal or recycling of this equipment and batteries. The main constituents of each part are listed in the table below. As there is a risk of infection, do not recycle patient attachments such as cuffs, but dispose of them as instructed by your facility's procedures and applicable regulations.
  • Page 34: Specifications Specifications

    Specifications Specifications Main unit Measurement NIBP, PR Parameter Main unit: 120 x 190 x 110 (mm) 4.72 x 7.48 x 4.33 (inch) (W x H x D) Dimension AC adapter: 46 x 66 x 37 (mm) 1.81 x 2.60 x 1.46 (inch) (W x H x D) Main unit: Approx.
  • Page 35 Type: AA batteries, x4 Number of operation cycles when fully charged: 250 • Measurement conditions - New battery fully charged (AA high-performance Dry cell manganese) battery - Ambient temperature of 23°C (73.4°F) - Using M-size cuff - SYS120 / DIA80 / PR60 - One 5-minute cycle consisting of "cuff measurement time + wait time"...
  • Page 36 Non-Invasive Blood Pressure (NIBP) Measurement Pulse rate accuracy technology Measurement Dynamic Linear Deflation method method Pressure 0 to 300 mmHg display range Pressure display Within ±3 mmHg or 2 % accuracy NIBP SYS 60 to 250 mmHg measurement DIA 40 to 200 mmHg range Pulse rate 40 to 200 /min NIBP...
  • Page 37: Correct Disposal Of This Product

    This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices. This medical device manufactured by your local OMRON representative conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: •...
  • Page 38: Manufacturer's Declaration

    Manufacturer's Declaration The HBP-1100 is intended for use in the electromagnetic environment specified below. The customer or the user of the HBP-1100 should assure that it is used in such an environment. Electromagnetic Emissions (IEC60601-1-2) Emission Test Compliance Electromagnetic Environment The HBP-1100 uses RF energy only for internal functions.
  • Page 39 70 % U IEC 61000-4-11 mains interruptions, it is for 25 cycles for 25 cycles recommended that the HBP-1100 be powered : Rated voltage of <5 % U <5 % U from an uninterruptible power supply or for 5 sec.
  • Page 40 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HBP-1100 is used exceeds the applicable RF compliance level above, the HBP-1100 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the HBP-1100.
  • Page 41 Recommended separation distance between portable and mobile RF communications equipment and the HBP-1100 The HBP-1100 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HBP- 1100 can help prevent electromagnetic interference by maintaining a minimum...
  • Page 42: Optional Accessories Optional Medical Accessories

    Optional Accessories Optional Medical Accessories AC adapter GS CUFF XL GS CUFF L GS CUFF M 60240HW5SW HXA-GCUFF- HXA-GCUFF- HXA-GCUFF- XLLB GS CUFF S GS CUFF SS Instruction HXA-GCUFF- HXA-GCUFF- Manual SSLB...
  • Page 44 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com OMRON HEALTHACARE SINGAPORE PTE LTD 438A Alexandra Road, #05-05/08 Alexandra Technopark, Asia Pacific HQ Singapore 119967 www.omron-healthcare.com.sg OMRON (DALIAN) CO., LTD.

Table of Contents