Automatic Blood Pressure Monitor
HBP-9020
Instruction Manual
Thank you for purchasing the OMRON HBP-9020 unit.
Read all of the instructions in the manual before you operate the unit
and keep the manual near the unit at all times for future reference.

Advertisement

Table of Contents
loading

Summary of Contents for Omron HBP-9020

  • Page 1 Automatic Blood Pressure Monitor HBP-9020 Instruction Manual Thank you for purchasing the OMRON HBP-9020 unit. Read all of the instructions in the manual before you operate the unit and keep the manual near the unit at all times for future reference.
  • Page 3: Table Of Contents

    Contents Exemptions ......................2 Notes on Safety...................... 3 Features of the Product ..................7 Components of the Product / Options ..............8 Names and Functions of the Components............9 How to Install the Monitor ................... 11 Place the monitor on the desk ................. 11 Attach the armrest....................
  • Page 4: Exemptions

    Exemptions Disclaimer Our company assumes no responsibility for the following: 1.Failures, damage, or injuries due to maintenance or repair work performed by other than our company or a company that we specify. 2.Failures or damage to one of our products caused by a product of another manufacturer not delivered by us.
  • Page 5: Notes On Safety

    Do not use the monitor in a place where it may get wet, such as by pool. This may result in fire or electrical shock. Do not install any component or equipment that is not specified by OMRON on this monitor. This may cause fire or electrical shock.
  • Page 6 Notes on Safety Warning: Power Insert the power plug all the way into the outlet. Otherwise electric shock, short circuiting, or fire may result. When removing the power plug from the outlet, grasp and pull on the plug, not the cord. Pulling on the cord may break wires and cause a short circuit, resulting in fire or electric shock.
  • Page 7 Notes on Safety Caution: General rules A person with a serious blockage of blood flow or other blood disorder should use this monitor only under the supervision of a doctor. Otherwise, compression by the monitor may cause temporary internal bleeding. If the monitor does not stop inflation during measurement or otherwise operates abnormally, press the [EMERGENCY STOP] switch.
  • Page 8 Notes on Safety General Advice • Do not place anything on this monitor. • Be sure to inspect the monitor before starting and perform regular maintenance inspections to ensure safe and proper use of the monitor. • If a measurement error occurs or a result is questionable, use auscultation or other method to check. - Measured values may be indicated erroneously or measurement may not be possible for a person who has significant short-term fluctuation of blood pressure such as an irregular pulse.
  • Page 9: Features Of The Product

    2. Turn off the power and unplug the monitor. 3. Place an "Out of Service" note on the monitor so nobody will use it. 4. Please contact your local authorized OMRON dealer/supplier. Features of the Product • Supports accurate measurement and can easily be used by a single person.
  • Page 10: Components Of The Product / Options

    Components of the Product / Options Before using this product, make sure that no accessories are missing and that neither the unit nor the accessories are damaged. Please contact your local authorized OMRON dealer/supplier if an accessory is missing or damaged.
  • Page 11: Names And Functions Of The Components

    Names and Functions of the Components Main Unit Armrest Place the arm on this armrest during measurement. [EMERGENCY STOP] switch Display Press to turn off the power Cuff in an emergency. Cuff Cover [START/STOP] switch (R) Press to start or [MODE] switch stop measurement.
  • Page 12 Names and Functions of the Components Rear Elbow rest The elbow is placed here before measurement starts. RS-232C connector Enables connection to a computer. SD memory card connector An SD memory card can be inserted in this connector. [PRINT] switch Selects whether or not results are printed.
  • Page 13: How To Install The Monitor

    How to Install the Monitor When installing the monitor, it is also recommended that the optional stand and chair be installed. (Page 8) If the optional stand and chair are not used, correct blood pressure measurement may not be possible. * If the optional stand and chair cannot be installed, use the following desk and chair: •...
  • Page 14: Attach The Instruction Panel And The Elbow Rest Guide Plate

    How to Install the Monitor Attach the instruction panel and the elbow rest guide plate Instruction panel Elbow rest guide plate How to connect to the power source Connect the provided power cord. Insert the socket side of the power cord into the AC connector on the back of the monitor.
  • Page 15: How To Set The Date And Time

    If the date appears when the power is turned on, set the date and time. If the same display appears the next time the power is turned on, the backup battery must be replaced. Please contact your local authorized OMRON dealer/supplier.
  • Page 16: How To Load Printer Paper

    • Do not load printer paper during initialization, measurement, or printing. This may cause a paper jam. • Replacement printer paper is available as an option. For details, please contact your local authorized OMRON dealer/supplier. Make sure that the [POWER]...
  • Page 17 How to Load Printer Paper Insert the printer paper in the orientation shown, and pull out about 10 cm. Make sure the printer paper is in the correct orientation. Close the paper cover. The printer paper will be cut automatically. Remove the cut paper.
  • Page 18: Checking The Unit Before Usage

    Checking the Unit Before Usage Before turning on the power, check the followings : External appearance • The monitor is not deformed due to falling or other impact. • The monitor is not dirty or wet. • The power cord is not damaged and there is no loose connections. •...
  • Page 19: How To Measure Blood Pressure

    How to Measure Blood Pressure While blood pressure is being measured, constantly monitor the user and the blood pressure monitor to make sure there are no problems. To print measurement results, printer paper must be loaded. (Page 14) Notes: • The volume of the audio voice and other settings related to measurement can be configured in function selection mode.
  • Page 20 How to Measure Blood Pressure Press the [START/STOP] switch. The cuff automatically wraps around the arm and the monitor starts measurement. The pressure will appear in the diastolic blood pressure display during measurement. • When the left arm is used for measurement, use the [START/STOP] switch on the right side of the monitor.
  • Page 21 How to Measure Blood Pressure Canceling measurement ■ If pain is felt in the arm or discomfort is experienced, remain calm and press the [START/STOP] switch again. Emergency stop ■ If a problem is encountered, for example, if the air in the cuff does not deflate even after the [START/STOP] switch is pressed, press the [EMERGENCY STOP] switch.
  • Page 22 How to Measure Blood Pressure Checklist for situations when the measurement results are doubtful or measurement is not possible 1. Basic points to ensure correct measurement The elbow detection lamp should be on during measurement. (If the elbow detection lamp is on during measurement, "Measurement was performed at the correct arm position"...
  • Page 23: Printing The Measurement Data

    "Measurement was performed at the correct *For a compatible SD memory card, please contact your local authorized OMRON dealer/supplier. arm position" will be indicated on the printed measurement results. Print pattern: A If the elbow detection lamp was still blinking when measurement was started, "Placing...
  • Page 24 Printing the Measurement Data Print pattern: B Print pattern: C...
  • Page 25: Function Selection Mode

    Function Selection Mode Function selection mode allows various settings and special functions to be used. (The settings are retained even if an interruption in the power supply occurs.) How to enter function selection mode and change settings To enter function selection mode, hold the [MODE] switch and then the [START/STOP] switch for at least 3 seconds until a beep sounds.
  • Page 26 Function Selection Mode Settings in function selection mode Factory Item Settings Description default 12/24 H 12 H, 24 H 24 H Select the time print format. selection 0, 1, 2, 3, 4, 5, Set the volume of the audio voice for the Volume 6, 7, 8, 9, 10 measurement results.
  • Page 27: Maintenance And Cleaning

    • If the cuff cover is dirty, remove it and wash by hand. • A replacement cuff cover is available as an option. For details, please contact your local authorized OMRON dealer/supplier. How to remove the cuff cover Make sure that the [POWER] switch is in the OFF position.
  • Page 28 Maintenance and Cleaning How to attach the cuff cover Insert the ring marked "UP" from the front as shown, and pull to the back. Insert the front of the cuff cover into the groove. Insert the back of the cuff cover into the groove.
  • Page 29: Cleaning Surface

    Disposal Cleaning Surface If the monitor is dirty, use a soft cloth moistened with diluted neutral detergent or diluted disinfecting alcohol (well wrung) to wipe off the dirt. Note, however, that the power connector should not be wiped or moistened in any way. Disposal Description Disposing or recycling this device could contaminate the environment.
  • Page 30: Errors Indicating Unsuccessful Measurement

    Accessory function errors These error codes appear in the systolic blood pressure display area. Error code Cause Remedial action Please contact your local authorized OMRON dealer/supplier. • Printer failure. (This error does not appear when the [PRINT] switch is set to "OFF".) •...
  • Page 31 • A failure has occurred in the 3. Place an "Out of Service" note on the monitor so monitor. that nobody will use it. 4. Please contact your local authorized OMRON dealer/supplier. Printer errors These error codes appear in the pulse rate display.
  • Page 32: Troubleshooting

    Troubleshooting If a problem occurs while using the monitor, check the following troubleshooting list. Problem Probable cause Remedial action Insert the power cord into the electric outlet Nothing appears on the The power cord is loose or correctly. display when the disconnected.
  • Page 33 * If the problem persists even after the above list has been checked, a failure may have occurred. For product failures and repairs, please contact your local authorized OMRON dealer/supplier. * In rare cases the body constitution of some people causes an error to occur and measurement is not possible. In this situation, please...
  • Page 34: Specifications

    Asia Pacific HQ 438A Alexandra Road, #05-05/08, Alexandra Technopark, Singapore 119967 Note: Specifications may be changed without prior notice. • This device supports SD Association standard SD memory cards. • OMRON HEALTHCARE Co., is a member of the SD Association.
  • Page 35: Manufacture's Declaration

    HBP-9020. The HBP-9020 is suitable for use in the specified electromagnetic environment. The customer and/ or user of the HBP-9020 should assure that it is used in an electromagnetic environment as described below; Electromagnetic Emissions (IEC60601-1-2)
  • Page 36 (60 % dip in UT) short interruptions environment. for 5 cycles for 5 cycles and voltage If the user of the HBP-9020 requires variations on continued operation during power mains 70 % UT 70 % UT power supply interruption, it is recommended that the...
  • Page 37 If the measured field strength in the location in which the HBP-9020 is used exceeds the applicable RF compliance level above, the HBP-9020 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the HBP-9020.
  • Page 38 Recommended separation distance between portable and mobile RF communications equipment and the HBP-9020 The HBP-9020 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HBP-9020 can help prevent electromagnetic interference...
  • Page 40 OMRON HEALTHCARE Co., Ltd. Manufacturer 24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto, 615-0084 Japan OMRON HEALTHCARE SINGAPORE PTE LTD. Asia Pacific HQ 438A Alexandra Road, #05-05/08, Alexandra Technopark, Singapore 119967 5329319-5D...