6

Operational Faults

6.1
Operational faults
Please proceed as follows if you have any problems:
· First assess the possible cause of the problem using the following table.
· Check the drive console status display. Evaluate the blink error code.
· Carry out the necessary checks and repairs as recommended in the following table.
6.1.1

Drive fault diagnosis

PROBLEM
Scooter will
not start
OTHER
POSSIBLE
SYMPTOMS
CAUSE
Status display
Batteries possibly
on operating
defective
console does
not illuminate
Batteries possibly
over-discharged
Power supply to
operating console
possibly
interrupted
Operating
console possibly
defective
Operating unit
Various causes
status display
blinking
SERVICE MANUAL
SOLUTION
Documentation
· Replace
See operating manual
batteries
· Pre-charge
See operating manual
batteries
· Check master
See operating
fuse
instructions for main fuse
position
· Check cable
See "Replacing the
between
electronic module" on
modules for
page 51
loose
connections or
damage
· Replace
See "Replacing
operating
operating console" on
console
page 38
· Assess error
See "Error Codes and
code
Diagnostic Codes" on
page 18