Invacare Orion Service Manual page 22

Hide thumbs Also See for Orion:
Table of Contents

Advertisement

SERVICE MANUAL
·
Remove both screws (2) from the wheel
cover (1).
·
Remove wheel cover (1).
·
Remove both screws (1) from the drive
holder (2).
·
Remove drive holder (2).
·
Remove screws (1) on disengaging lever
(2).
·
Remove disengaging lever (2).
·
Lifted drive out of its holder and take it out of
the frames to the side.
·
Replace drive unit.
·
Reassembly of the drive takes place in
reverse order
·
Check all functions (trial run).
22
®